41902-100 | Quote |
---|
Excellent for the routine maintenance of image quality in ultrasonography as well as the ageing of devices and transducers.
US-4B
41902-100 | Quote |
---|
Excellent for the routine maintenance of image quality in ultrasonography as well as the ageing of devices and transducers.
Features | 1. Compact size allows imaging in a single scan, saving time and si0mplifying the process. 2. To ensure the verticality of the entrance beam, two horizontal wires are embedded and the supporting guide frame secure the transducer in its home position for testing, 3. For monthly basic quality test, as well as longer term quality assurance to maintain consistency of the performance of scanners and transducers. 4.Stable and long-life material of the phantom *proved by 5-year-test 5.To maintain the internal temperature of the phantom (23°C to 26°C), a thermometer is provided. |
---|---|
Training skills / Applications | Image quality and aging of the transducer |
Set includes | 1 phantom, 1 thermometer, 1 Instruction manual |
Size (approx.) | W10.3 x D5 x H8cm/W4 x D1.97 x H3.15in. |
Packing size (approx.) | W39 x D28 x H21cm / W15.35 x D11 x H8.27in. |
Weight (approx.) | 0.33kg/ 0.72 lbs |
Packing weight (approx.) | 3kg/ 6.6lbs |
Materials | Urethane elastomer, acryl, nylon Latex free |
Production & Development Supervision | The Japan Association of Breast and Thyroid Sonology (JABTS) |
Material features | Sonic velocity: 1434(±1)m/ sec at 25°C Attention coefficient: 0.59 dB/ cm MHz at 25°C Acoustic impedance: 1.37 rayl at 25°C |
Update | January 27, 2023 |